Specifications |
Channels |
0.3 μm, 0.5 μm, 1.0 μm, 2.0 μm, 5.0 μm, 10.0 μm |
Flow rate |
0.1 cfm (2.83 L/min) |
Light source |
775 nm to 795 nm, 90 mW class 3B laser |
Count modes |
Raw counts, #/m3, #/ft3, #/liter in Cumulative or Differential mode |
Counting efficiency |
50% @ 0.3 μm; 100% for particles > 0.45 μm (per ISO 21501) |
Zero count |
1 count / 5 minute (JIS B9921) |
Concentration limits |
10% at 4,000,000 particles per ft3 (per ISO 21501) |
Data storage |
10,000 records (rotating buffer) |
Alarms |
Counts, low battery, sensor fail |
Delay time |
0 to 24 hours |
Sample inlet |
Isokinetic probe |
Security |
Administrator password control (optional) |
Communication mode |
USB or Etheret |
Dimensions (L x W x D) |
27.2 x 9.9 x 5.3 cm (10.7 x 3.9 x 2.1 in) |
Weight |
0.68 kg (1.5 lb) |
Power adapter |
100 V AC to 240 V AC, 12 V DC, 2.5 A |
Rechargeable battery |
Li-ion 7.4 V 2600 mAh; 10 hours typical use (5 hours continuous sampling) / 3.5 hours |
Calibration |
PSL particles in air (NIST traceable) |
Operating environment |
50°F to 104°F (10°C to 40°C) / < 95% non-condensing relative humidity |
Storage environment |
14°F to 122°F (-10°C to 50°C)/Up to 98% non-condensing relative humidity |
Warranty |
One-year |
Accessories included |
Cradle for charging and convenient USB and Ethernet communications, ENET CAT5E 7 foot cable, USB-A to MINI-B 6 foot cable, 12 V dc power supply, zero count inlet- filter, filter adapter, sample inlet protective cap, hard case, getting started manual and users manual CD |